Chemistry and hematology: common CMS-2567 deficiency patterns

In chemistry and hematology, the tags cited most often are about documented conditions and materials: storage and operating conditions (D5413), expired or deteriorated reagents (D5417), instrument maintenance (D5429) and verifying performance before patient testing (D5421). Calibration and QC review findings are also common in the analytic phase.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Chemistry and hematology CLIA 2567 deficiency examples
Chemistry and hematology CLIA 2567 deficiency examples

Examples on this site are illustrative composites written to show how findings are typically structured. They are not quotations from any specific laboratory's CMS-2567.

D5413 · 42 CFR 493.1252(b)

Typical finding pattern: Reagent refrigerator temperatures were not recorded on multiple days, or the acceptable range was not defined, and no corrective action was documented for out-of-range readings.

What a complete correction covers: Define acceptable ranges consistent with manufacturer instructions; review whether reagents stored out of range were used and assess affected results; add daily review with escalation; monitor logs weekly for a set period and report to the director.

D5417 · 42 CFR 493.1252(d)

Typical finding pattern: Control or reagent lots past their expiration date were found in use on an analyzer.

What a complete correction covers: Remove and discard; identify patient runs performed with expired material and evaluate need for corrected reports; institute receiving and open-date labeling; add a monthly expiration sweep with sign-off.

D5421 · 42 CFR 493.1253(b)(1)

Typical finding pattern: A new FDA-cleared analyzer was placed into patient use without documented verification of accuracy, precision, reportable range and reference intervals.

What a complete correction covers: Complete and document verification; evaluate results reported before verification; require director approval of verification before go-live on every new test system.

D5429 · 42 CFR 493.1254(a)(1)

Typical finding pattern: Daily and weekly maintenance specified by the manufacturer was not documented for an extended period.

What a complete correction covers: Restore maintenance; review whether missed maintenance affected results; build maintenance into the shift checklist; supervisor reviews completion weekly.

Calibration and calibration verification

Under §493.1255, calibration and calibration verification must follow the manufacturer's instructions and occur at least every six months for many test systems, and whenever certain events occur (reagent lot changes affecting results, major maintenance, QC trends). Findings typically note a missing six-month verification or no documentation after a major repair.

Quality control review

§493.1256 sets QC frequency and requires corrective action when controls fail. Write-ups often show controls out of range with patient results released and no documented investigation. A complete answer addresses the patient results reported during the failure, not just the control.

Hematology specifics

Hematology citations frequently involve manual differential competency, slide review criteria in the procedure manual, and documentation of analyzer flags and delta checks. Tie the correction to the written procedure (D5401/D5403) so the fix outlives the people who made it.

Want a second set of eyes on your 2567?

Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What is D5413 in a chemistry lab?

A citation under 42 CFR 493.1252(b) when the lab has not defined, monitored or documented the conditions essential for reagent and specimen storage, test system operation and result reporting, such as temperature and humidity.

How often is calibration verification required?

Under 42 CFR 493.1255, at least once every six months for many test systems and when specified events occur, following the manufacturer's instructions.

What should a plan of correction for expired reagents include?

Removal of the material, review of patient results run with it, a receiving and labeling process to prevent recurrence, and ongoing monitoring with sign-off.