Microbiology and molecular: common CMS-2567 deficiency patterns
Microbiology and molecular write-ups often focus on whether procedures exist and are followed (D5401, D5403), whether media, reagents and controls are checked and in date (D5417 and specialty QC under 42 CFR 493.1261–493.1265), whether new methods were verified or validated (D5421), and whether patient reports carry required elements (D5805).

Examples on this site are illustrative composites written to show how findings are typically structured. They are not quotations from any specific laboratory's CMS-2567.
D5401 / D5403 · 42 CFR 493.1251
Typical finding pattern: The written procedure for a PCR assay did not match the steps staff performed, or lacked specimen acceptability and rejection criteria.
What a complete correction covers: Revise the procedure to match the validated method, with director approval and dating; retrain staff with documented review; add annual procedure review.
Specialty QC · 42 CFR 493.1261–493.1265
Typical finding pattern: Media QC or identification system QC was not performed or documented as required for the specialty.
What a complete correction covers: Perform and document required QC; evaluate results reported without it; add QC to the bench worksheet with supervisor review.
D5421 · 42 CFR 493.1253
Typical finding pattern: A laboratory-modified molecular assay was placed in use without documentation establishing performance specifications.
What a complete correction covers: Complete validation for modified or laboratory-developed tests under §493.1253(b)(2); hold or qualify results; require director sign-off before patient use.
D5805 · 42 CFR 493.1291(c)
Typical finding pattern: Patient reports did not show the name and address of the testing laboratory, or lacked required specimen information.
What a complete correction covers: Correct the report template in the LIS; review and correct affected reports; verify report content at each LIS change.
Contamination and workflow
Molecular labs are often asked how they prevent amplicon contamination: unidirectional workflow, separate pre- and post-amplification areas, and monitoring with no-template controls. If the procedure manual describes controls the bench does not run, expect a D5401 finding.
Verification vs validation
Unmodified FDA-cleared tests need verification of the manufacturer's performance specifications (§493.1253(b)(1)). Modified or laboratory-developed tests need the lab to establish performance specifications (§493.1253(b)(2)). Mixing up the two is a common root cause in molecular findings.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
What CLIA regulation covers microbiology QC?
Specialty and subspecialty QC requirements for microbiology are in 42 CFR 493.1261 through 493.1265, alongside the general QC rules in 493.1256.
Do laboratory-developed tests need validation under CLIA?
Yes. Under 42 CFR 493.1253(b)(2), a laboratory that modifies an FDA-cleared test or uses a test not subject to FDA clearance must establish performance specifications before reporting patient results.
What does D5805 cover?
Required test report content under 42 CFR 493.1291(c), including patient identification and the name and address of the testing laboratory.