Waived and PPM labs: common CLIA deficiency patterns

Waived labs must follow the manufacturer's instructions and test only within the scope of their certificate (42 CFR 493.15 and 493.35–493.39). The most common waived findings are not following package inserts, testing beyond the certificate, and missing records the instructions require. PPM labs add the provider-performed microscopy rules.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Waived and PPM CLIA deficiency examples for certificate of waiver labs
Waived and PPM CLIA deficiency examples for certificate of waiver labs

Examples on this site are illustrative composites written to show how findings are typically structured. They are not quotations from any specific laboratory's CMS-2567.

Waived (D1000 series) · 42 CFR 493.15(e)

Typical finding pattern: Staff did not follow the manufacturer's instructions, for example running controls less often than the package insert requires or using test kits past their dating.

What a complete correction covers: Retrain using the current insert; evaluate results produced outside instructions; post a checklist from the insert; supervisor spot-checks with documentation.

Scope · 42 CFR 493.35–493.39

Typical finding pattern: The site performed a test not categorized as waived while holding only a Certificate of Waiver.

What a complete correction covers: Stop the test immediately; evaluate reported results; apply for the correct certificate on Form CMS-116 before resuming; verify categorization of every test on the menu.

PPM · 42 CFR 493.19

Typical finding pattern: Microscopy procedures were performed by a person not authorized under the PPM certificate, or PPM procedures were not documented under the quality requirements that apply to moderate complexity testing.

What a complete correction covers: Restrict PPM procedures to qualified providers; document competency and QC as required; review the PPM list on the certificate.

How waived labs are surveyed

Certificate of Waiver labs are not routinely surveyed every two years like Certificate of Compliance labs, but CMS and state agencies do visit a share of them and respond to complaints. When problems are found, the CMS-2567 process applies the same way.

Upgrading to moderate complexity

If a site needs a test that is not waived, it must apply for the right certificate on Form CMS-116 and meet the personnel and quality system requirements that come with it before testing. See CLIA certificate types.

Want a second set of eyes on your 2567?

Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Can a waived lab get a CMS-2567?

Yes. If a survey of a Certificate of Waiver lab finds noncompliance, the findings are recorded on Form CMS-2567.

What is the most common waived testing deficiency?

Not following the manufacturer's instructions, which 42 CFR 493.15(e) requires for waived tests.

What is a PPM certificate?

A Certificate for Provider-Performed Microscopy Procedures, which allows physicians, midlevel practitioners or dentists to perform specified microscopy procedures, plus waived tests.