CLIA 2567s LibraryCMS-2567 deficiency patterns by specialty Book a free consultation
Home › Personnel & PT

Personnel and proficiency testing: common CMS-2567 patterns

Personnel and proficiency testing findings dominate CMS's condition-level list. In 2024, D2016 (unsuccessful PT participation) was cited at 738 labs, D6000 (moderate complexity director) at 709, and D6076 (high complexity director) at 388. At standard level, D5209 competency policies (778 labs) and D6016 director PT responsibilities (451 labs) are among the most cited.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Personnel and proficiency testing CLIA 2567 deficiency examples
Personnel and proficiency testing CLIA 2567 deficiency examples

Examples on this site are illustrative composites written to show how findings are typically structured. They are not quotations from any specific laboratory's CMS-2567.

D2016 · 42 CFR 493.803

Typical finding pattern: The lab had unsatisfactory results for the same analyte in two consecutive or two of three PT events, making participation unsuccessful.

What a complete correction covers: Investigate each failed event, review patient results for the period, correct the root cause, document director review, and follow any CMS requirements on ceasing testing for the analyte.

D6000 / D6076 · 42 CFR 493.1403 / 493.1441

Typical finding pattern: The director did not fulfill responsibilities in §493.1407 or §493.1445, for example no evidence of oversight of QC, PT or competency.

What a complete correction covers: Document director oversight activities, onsite visits and reviews; correct each underlying deficiency; establish a recurring director review calendar.

D5209 · 42 CFR 493.1235

Typical finding pattern: Written competency policies were missing, or six-element competency was not completed for testing personnel at six months and annually.

What a complete correction covers: Write or revise the competency policy; complete overdue assessments with all six elements; track due dates with supervisor review.

D6033 · 42 CFR 493.1409

Typical finding pattern: No technical consultant meeting §493.1411 qualifications was in place for moderate complexity testing.

What a complete correction covers: Appoint a qualified technical consultant with documented qualifications and delegated duties; review work done without one.

The six elements of competency

For nonwaived testing, competency assessment includes direct observation of testing, monitoring result recording and reporting, review of worksheets and QC records, direct observation of maintenance, assessment of test performance through blind or previously tested samples, and assessment of problem-solving skills (§493.1451, §493.1413).

Never refer PT samples

Sending PT samples to another lab, or receiving another lab's PT samples, is prohibited under §493.801(b)(4) and can lead to certificate revocation. If a sample was referred, treat it as the most urgent item on the 2567.

Want a second set of eyes on your 2567?

Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What is D2016?

The condition-level tag for successful participation in proficiency testing under 42 CFR 493.803. It is the most-cited CLIA condition in CMS's 2024 data.

What does D6000 mean?

The condition for a moderate complexity laboratory director who meets 42 CFR 493.1405 qualifications and provides overall management under 493.1407.

How often is competency assessed?

For nonwaived testing, semiannually during the first year an individual tests patient specimens, then annually, using six elements.