Responding to a Alcohol (Blood Ethanol) Deficiency on the CMS-2567
A strong plan of correction for a Blood Alcohol finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Blood Alcohol records that prove it attached.
What a Blood Alcohol plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Blood Alcohol and when.
- The look-back: how you found other Blood Alcohol results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Blood Alcohol
- Daily Blood Alcohol QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month Blood Alcohol calibration verification and reportable range data with director review
- Blood Alcohol PT enrollment, attestation statements, graded reports, investigation of every miss
- Blood Alcohol verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Blood Alcohol reference intervals
- Training and competency records for everyone who runs Blood Alcohol
Blood Alcohol PT and accuracy
Alcohol (Blood Ethanol) is a regulated analyte (toxicology, 42 CFR 493.937). CLIA PT acceptance limit: target Value ±20%. If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Blood Alcohol findings
- Do not clean the site with alcohol swabs; use a non-alcohol antiseptic.
- Keep tubes capped; ethanol evaporates from open containers.
- Medical serum/plasma results are not the same as whole blood forensic results; state specimen type.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Acetaminophen (Serum) (full guide)
- Salicylate (full guide)
- Urine Drug Screen (Drugs of Abuse) (full guide)
Full guide: Alcohol (Blood Ethanol) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Blood Alcohol deficiency?
State the correction, the look-back for affected Blood Alcohol results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Blood Alcohol plan of correction?
Daily Blood Alcohol QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month Blood Alcohol calibration verification and reportable range data with director review; Blood Alcohol PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.937, Toxicology
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)