Responding to a Blood Gases (pH, pO2, pCO2) Deficiency on the CMS-2567
A strong plan of correction for a Blood Gas finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Blood Gas records that prove it attached.
What a Blood Gas plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Blood Gas and when.
- The look-back: how you found other Blood Gas results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Blood Gas
- Daily Blood Gas QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month Blood Gas calibration verification and reportable range data with director review
- Blood Gas PT enrollment, attestation statements, graded reports, investigation of every miss
- Blood Gas verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Blood Gas reference intervals
- Training and competency records for everyone who runs Blood Gas
Blood Gas PT and accuracy
Blood Gases (pH, pO2, pCO2) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: pH: target value ±0.04; pCO2: target value ±8% or ±5 mm Hg (greater); pO2: target value ±15% or ±15 mmHg (greater). If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Blood Gas findings
- Air bubbles left in the syringe shift pO2 and pCO2; expel air and cap immediately.
- Delayed analysis lets cells keep metabolizing; follow your analysis-time and storage rules.
- Incorrect heparin amount or liquid heparin dilution affects results; use the collection devices your procedure specifies.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Potassium (full guide)
- Sodium (full guide)
- Ionized Calcium (full guide)
- Lactate (Lactic Acid) (full guide)
Full guide: Blood Gases (pH, pO2, pCO2) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Blood Gas deficiency?
State the correction, the look-back for affected Blood Gas results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Blood Gas plan of correction?
Daily Blood Gas QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month Blood Gas calibration verification and reportable range data with director review; Blood Gas PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.931, Routine chemistry
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)