Responding to a Complete Blood Count (CBC) Deficiency on the CMS-2567
A strong plan of correction for a CBC finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the CBC records that prove it attached.
What a CBC plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for CBC and when.
- The look-back: how you found other CBC results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for CBC
- Daily CBC QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month CBC calibration verification and reportable range data with director review
- Twice-yearly CBC accuracy checks (external PT, split samples or reference material) with acceptance criteria
- CBC verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing CBC reference intervals
- Training and competency records for everyone who runs CBC
CBC PT and accuracy
Not listed in Subpart I: no required PT, but accuracy must be verified at least twice a year (42 CFR 493.1236(c)). If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind CBC findings
- The CBC is a panel: WBC, RBC, hemoglobin, hematocrit, platelet count and differential each carry their own PT limits under 42 CFR 493.941.
- Define smear review criteria for analyzer flags and delta checks.
- Verify carryover, linearity and reference intervals for each analyzer.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Hemoglobin (full guide)
- Leukocyte Count (WBC)
- Platelet Count (full guide)
- White Blood Cell Differential (full guide)
Full guide: Complete Blood Count (CBC) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a CBC deficiency?
State the correction, the look-back for affected CBC results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a CBC plan of correction?
Daily CBC QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month CBC calibration verification and reportable range data with director review; Twice-yearly CBC accuracy checks (external PT, split samples or reference material) with acceptance criteria.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)