Responding to a Ferritin Deficiency on the CMS-2567
A strong plan of correction for a Ferritin finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Ferritin records that prove it attached.
What a Ferritin plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Ferritin and when.
- The look-back: how you found other Ferritin results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Ferritin
- Daily Ferritin QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month Ferritin calibration verification and reportable range data with director review
- Ferritin PT enrollment, attestation statements, graded reports, investigation of every miss
- Ferritin verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Ferritin reference intervals
- Training and competency records for everyone who runs Ferritin
Ferritin PT and accuracy
Ferritin is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±20%. If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Ferritin findings
- Ferritin rises with inflammation; that is a clinical factor, so do not adjust reference intervals for it.
- Very high results may need dilution; verify the dilution protocol as part of the reportable range.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Iron, Total (full guide)
- TIBC (Total Iron-Binding Capacity), Direct (full guide)
- Hemoglobin (full guide)
Full guide: Ferritin testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Ferritin deficiency?
State the correction, the look-back for affected Ferritin results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Ferritin plan of correction?
Daily Ferritin QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month Ferritin calibration verification and reportable range data with director review; Ferritin PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.931, Routine chemistry
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)