Responding to a D (Rh) Typing Deficiency on the CMS-2567
A strong plan of correction for a Rh (D) Typing finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the Rh (D) Typing records that prove it attached.
What a Rh (D) Typing plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for Rh (D) Typing and when.
- The look-back: how you found other Rh (D) Typing results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for Rh (D) Typing
- Daily Rh (D) Typing QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Rh (D) Typing PT enrollment, attestation statements, graded reports, investigation of every miss
- Rh (D) Typing verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing Rh (D) Typing reference intervals
- Training and competency records for everyone who runs Rh (D) Typing
Rh (D) Typing PT and accuracy
D (Rh) Typing is a regulated analyte (immunohematology, 42 CFR 493.959). CLIA PT acceptance limit: 100% accuracy. If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind Rh (D) Typing findings
- Positive patient identification at collection is the single most important control; use your two-identifier and labeling policy.
- ABO group must be determined with at least anti-A and anti-B reagents and confirmed by testing the serum with known A1 and B red cells (42 CFR 493.1271(a)(2)).
- Weak D testing policy should be written and followed consistently.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
Full guide: D (Rh) Typing testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a Rh (D) Typing deficiency?
State the correction, the look-back for affected Rh (D) Typing results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a Rh (D) Typing plan of correction?
Daily Rh (D) Typing QC records, Levey-Jennings charts, corrective action log, control lot ranges; Rh (D) Typing PT enrollment, attestation statements, graded reports, investigation of every miss; Rh (D) Typing verification study: accuracy, precision, reportable range, reference interval check, director sign-off.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.959, Immunohematology
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)